OMD-525X – Stationary oxygen analyzer with IP66 enclosure
The OMD-525X from Southland Sensing is a stationary oxygen analyzer designed for continuous and accurate measurement of oxygen (O₂) in industrial processes and gas applications. The instrument combines electrochemical sensor technology with a robust design, high measurement accuracy, and low maintenance requirements.
The analyzer is equipped with an IP66/NEMA 4X-rated aluminum enclosure for wall mounting, enabling installation in both indoor and outdoor environments.
The OMD-525X can be configured for the measurement of very low oxygen concentrations at ppm levels, percentages, or for controlling oxygen purity. Depending on the selected sensor, the instrument offers measurements from 0.01 ppm up to 100% O₂.
The large backlit LCD display and user-friendly menu system make the instrument easy to configure and use. Integrated data logging via USB enables documentation and monitoring of the process oxygen level over time.
The analyzer is equipped with analog and digital outputs as well as two configurable alarm relays for easy integration with parent control and monitoring systems.
Features and benefits
- Continuous measurement of oxygen (O₂)
- Measurement from 0.01 ppm up to 100% O₂, depending on the sensor
- Electrochemical sensor technology with high precision
- IP66/NEMA 4X-rated enclosure
- Adapted for indoor and outdoor installation
- Integrated data logging via USB
- Two configurable alarm relays with time delay
- Analog output signal 4–20 mA
- Digital communication via Modbus RS485 ASCII
- Integrated temperature compensation
- Large backlit LCD display
- Low maintenance requirements
- Possibility of a complete test gas system
Measurement technology
The OMD-525X uses electrochemical oxygen gas sensors developed and manufactured by Southland Sensing. The sensors offer high accuracy, good measurement stability, and long lifespan without the need for electrolyte replenishment or electrode cleaning.
The instrument can be configured with various sensors to adapt to the measurement range and the composition of the sample gas. The standard sensors are suitable for applications such as nitrogen, hydrogen, helium, and hydrocarbons, while specially adapted sensors are available for applications involving carbon dioxide (CO₂) and natural gas.
Integrated data logging
The OMD-525X has integrated data logging where measurement values are stored directly on a removable USB memory in CSV format, compatible with Microsoft Excel.
The logging interval can be configured between 1 and 120 minutes, providing good opportunities for process monitoring, quality control and historical documentation.
Areas of application
- Process monitoring in industry
- Choking generators
- Oxygen generators
- PSA system
- Carbon dioxide (CO₂) breathalyser
- The food and beverage industry
- Checking the purity of the gas
- Inertization and protective gas atmospheres
- Industrial welding
- Additive manufacturing and 3D printing
- Air separation plants
- Laboratories and research facilities
Technical specifications
| Parameter | Technical specification |
|---|---|
| Measuring gas | Oxygen gas (O₂) |
| Measurement principle | Galvanic electrochemical sensor |
| Measuring ranges, ppm | 0–10, 0–100, 0–1 000 and 0–10 000 ppm |
| Measuring ranges, percent | 0–1, 0–5, 0–10, 0–25 and 0–100% |
| Accuracy | ±1% of full measurement range under constant conditions |
| Protection class | IP66/NEMA 4X |
| Display | Backlit LCD |
| Analog output signal | 4–20 mA |
| Digital communication | Modbus RS485 ASCII |
| Emergency exits | 2 configurable switching relays |
| Data logging | USB memory, CSV format |
| Logging interval | 1–120 minutes |
| Sample gas flow | Approximately 0.24–2.36 standard liters/minute |
| Gastric pressure | 0.1–50 PSIG |
| Operating temperature | 0 to +50 °C |
| Supply voltage | 100–240 VAC or 10–28 VDC, depending on the configuration |
| Gas connections | 1/8”, 1/4” or 6 mm |
| Enclosure material | Anodized aluminum |
| Dimensions | Approx. 241 × 165 × 97 mm |
| Weight | About 2 kg |
| Mounting | Wall mounting |
| Manufacturers | Southland Sensing, USA |
Test gas systems and options
The OMD-525X can be supplemented with ready-made sample gas systems to ensure the correct sample gas conditions up to the analyzer.
SS-1200 – Test gas system for high pressure and high flow
- Pressure regulator for incoming pressure up to approximately 41.4 barg (600 PSIG)
- Four-way bypass valve
- Manometer
- Flow meter
- Stainless steel mounting plate 304
SS-1400 – Test gas system for low pressure and low flow
- Integrated test gas pump
- Coalescence filter of 0.1 µm
- Stainless steel mounting plate 304
- Adapted for applications where the sample gas needs to be drawn up to the analyzer